Digital transducer electronis PAD4002A

Special features

  • Electronics suitable for full bridge strain gage transducers
  • Rugged connections, M12, 8-pin
  • Degree of protection up to IP68/IP69K, depending on which plug is used
  • Connection cables for sensor and digital output available as accessories
  • 2 freely programmable digital I/Os, e.g. for filling or monitoring applications
  • Can be combined with analog load cells or force transducers
  • The intuitive and user-friendly software PanelX is available free of charge for configuration, measurement and analysis

Type

PAD4002A−RS4

PAD4002A−CAN

Suitable for transducer types

Full bridge strain gages

Max. number of verification intervals per OIML R76,

d = e

6000

Class III, IIII (in preparation)

Multi-range applications (in preparation)

d = e

2 x 3000

Input sensitivity

μV/e

2::0.5

Legal-for-trade mode

industrial mode

μV/d

2::0.1

Measuring range

Resolution of nominal measuring range

mV/V digits

nominal ±2, max. ±3.2

5,120,000

Minimum transducer resistance

Ω

300

Maximum transducer resistance

Ω

1200

Transducer excitation voltage (carrier frequency 1.2 kHz)

VAC

5

Load cell connection

Maximum cable length to transducer 

m

4-wire circuit 6

Temperature coefficient of the zero signal per 10 K

%

±0.0055

Temperature coefficient of the sensitivity per 10 K

±0.0083

Non-linearity

% of meas. range

±0.0025

Voltage supply

V W A

+12 … +30, nominal 24 V

::;3

1.1

Supply voltage UB (DC)

Power consumption (350 Ω transducer resistance)

Max. current

Measurement signal resolution

bit

24

Data rate

1/s

4 … 1200

Digital filter bandwidth

Hz

0.1 … 120

Tare range (subtractive)

% of meas. range

+100

±100

Legal-for-trade mode industrial mode

Range of zero setting

% of meas. range

±2

±2

Legal-for-trade mode industrial mode

Interfaces

Max. number of bus nodes

90

CANopen interface Baud rate

Maximum cable length

baud m

Standard CiA DS301 10,000 … 1,000,000

::;5000 (10 kBaud) … ::;100 (500 kBaud) … ::;25 (1 MBaud)

RS−485 interface

baud

9600/19,200/38,400/57,600/115,200

Baud rate

Maximum cable length

m

50

Digital HCMOS input

V

0 … +12

Allowed input voltage

Low level

V

< 1

High level

V

> 4

Input resistance

70

Digital PLC input

V

0 … +30

Allowed input voltage

Low level

V

< 6

High level

V

> 10

Input resistance

9

Control outputs

V A A

11 … +30

< 0.5

< 1

External supply voltage

Max. current per output

Max. total current of all outputs

Digital transducer electronis PAD4002A - datasheet PDF, 0.17 Kb