KontrollVahend ja Tehnika (KVT) OÜ
Exclusive representative
Hottinger Baldwin Messtechnik GmbH, Germany
+372 6006190
+372 6006680


Products

Single Point Load Cell PW6D

Special features

  • Accuracy class C3 with OIML-R60 test report
  • Max. capacities: 3 kg ... 40 kg
  • Off center load compensated (OIML R 76)
  • Degree of protection IP67 (according to EN 60 529)
  • Shielded connection cable
  • Optimized for dynamic weighing applications
  • Optional:

  • Connection cable in six wire circuit
  • Different cable lengths
  • Aligned output, suitable for con- nection in parallel

Type

PW6D...

Accuracy class 1)

Maximum number of load cell intervals (nLC)

C3, C3MR

3000

Maximum capacity (Emax)

kg

3

5

10

15

20

30

40

Minimum LC verification interval (vmin),

(Accuracy class C3)

g

0.5

1

2

2

5

5

10

Temperature effect on zero balance (TK0), (Accuracy class C3)

% of Cn

/ 10 K

±0.0233

±0.0280

±0.0280

±0.0186

±0.0350

±0.0233

±0.0350

Minimum LC verification interval (vmin),

(Accuracy class C3MR)

g

0.2

0.5

1

1

2

2

5

Temperature effect on zero

balance (TK0), (Accuracy class C3MR)

% of Cn

/ 10 K

±0.0093

±0.0140

±0.0140

±0.0093

±0.0140

±0.0093

±0.0175

Max. platform size

mm

300 x 300

Sensitivity (Cn)

mV/V

2.0 ±0.2

Zero signal

0 ±0.1

Temperature effect on sensitivity (TKC) 2)

in the temperature range

+20 ... +40 °C [+68 ... +104 °F]

−10 ... +20 °C [+14 ... +68 °F]

% of Cn

/ 10 K

±0.0175

±0.0117

Relative reversibility error (dhy) 2)

% of Cn

±0.0166

Non-linearity (dlin) 2)

±0.0166

Ratio of minimum dead load output return (DR)

±0.0166

Off-center load error 3)

±0.0233

Input resistance (RLC)

380 ±38

Output resistance (R0)

380 ±38

Reference excitation voltage (Uref)

V

5

Nom. range of excitation voltage (BU)

1 ... 12

Isolation resistance (Ris) at 100 VDC

GΩ

> 2

Nominal (rated) range of ambient tempe- rature (BT)

°C [°F]

−10 ... +40 [+14 ... +104]

Operating temperature range (Btu)

−10 ... +50 [+14 ... +122]

Storage temperature range (Btl)

−25 ... +70 [−13 ... +158]

Limit load (EL) *)

*) at max. eccentricity

%

of Emax mm

150

100

Lateral load limit (Elq), static

%

of Emax

300

Breaking load (Ed)

300

Nominal (rated) displacement at Emax

(snom), approx.

mm

< 0.18

< 0.18

< 0.19

< 0.21

< 0.23

< 0.28

< 0.29

Natural frequency, approx.

Hz

270

390

500

600

675

760

790

Weight (G), approx.

kg

0.25

Degree of protection acc. to EN 60 529 (IEC 529)

IP67

Material: Measuring body

Application protection Cable sheath

Aluminum Silicone caoutchouc

PVC

Single Point Load Cell PW6D - datasheet PDF, 0.2 Kb